TRANSMITTER Model HR-LNC
Brand: TechnipFMCPenjelasan Produk :
Model HR-LNC
Smith Meter PD Meter Accessories
The Model HR-LNC (High Resolution – Large Numeral Conter) Transmitter is used to provide a digital signal for remote instrumentation.
The Transmitter output can either be single-channel with 100 pulses per revolution (of LNC right-hand wheel) or dual-channel, with 50 pulses per channel per revolution (of LNC right-hand wheel). -The dual channel pulse train is overlapping and channel1 leads chnnel 2 by 90°. This dual-channel overlapping pulse train can be used as a signal to a smith meter Comparator Totalizer (CDC-45 or CDC-41) which detects and alarms on; failure of a pulse transmission line; failure of a transmitter chip; or noise on the transmission lines.
|
Model No. |
Closures per Revolution of Right-hand Wheel |
|
HR-LNC-100 |
100 pulses (Single - Channel) |
|
HR-LNC-50 |
50 pulses (Dual-Channel) |
|
Specifications |
|
Input Power |
|
Normal Uperation: 10 to 15 Vdc; 65 mA max. Single channel; 100 mA dual channel |
|
Special Operation: 6 Vdc for DFR Indicator |
|
Output Voltage |
||||
|
Logic Off: 250 mV maximum |
|
|||
|
Logic On: |
|
|||
|
Input Impedance (W) |
100K |
22K |
1K |
330 |
|
output vs. Input Voltage (%) |
99 |
97 |
60 |
33 |
|
On Time: 50% ± 10%. Channel 1 leads Channel 2 by 90° ±18° |
|
Switching Time: |
|
Rise Time: 0.002 seconds maximum at 1 to 10 mA |
|
Fall Time: 0.0005 seconds maximum |
|
Input Shaft Speed: 250 rpm maximum |
|
Torque: 2 oz-in (1N.cm) maximum |
|
Operating temperature: -40°F to 180°F (-40°C to 82°C) |
|
Stroge Temperature: -67°F to 257°F (-55°C to 125°C) |
|
Approval: UL listed for hazardous locations: Class I, Group D, Division 1 |